The Molecular Devices SpectraMax 190 Microplate Reader is a high-performance UV-visible absorbance microplate reader designed for molecular biology laboratories, biotechnology research institutions, pharmaceutical companies, clinical research facilities, and academic laboratories.
Manufactured by Molecular Devices, the SpectraMax 190 is designed to provide accurate, reliable, and repeatable absorbance measurements for a wide range of laboratory applications. With a wavelength range of 190–850 nm, this instrument is particularly suitable for ultraviolet and visible light analysis, including DNA and RNA quantification, protein concentration determination, ELISA assays, and biochemical research.
The SpectraMax 190 features a monochromator-based optical system that allows researchers to select wavelengths in 1 nm increments. This provides considerable flexibility when performing different absorbance-based experiments. Its broad wavelength coverage makes it especially useful for laboratories that require both UV and visible absorbance measurements.
The instrument is optimized for 96-well microplate applications and supports endpoint and kinetic measurement modes. It can measure a complete 96-well microplate in approximately 12 seconds under standard endpoint reading conditions, helping laboratories improve sample processing efficiency.
Another important feature is its patented PathCheck technology. This technology automatically compensates for differences in liquid pathlength between microplate wells and standard 1 cm cuvettes. As a result, absorbance measurements can be converted into comparable cuvette-equivalent values, improving the consistency of quantitative analysis.
With its compact design, reliable optical performance, and compatibility with SoftMax Pro software, the SpectraMax 190 is a practical solution for laboratories performing routine UV-visible microplate analysis.
The SpectraMax 190 offers a wavelength range of 190–850 nm, covering both ultraviolet and visible light regions.
This capability enables researchers to perform nucleic acid measurements at 260 nm, protein analysis at 280 nm, and colorimetric assays at commonly used visible wavelengths.
The monochromator allows wavelength selection in 1 nm increments, providing flexibility for different experimental protocols.
The instrument uses a xenon flash lamp as its light source and a monochromator for wavelength selection.
Its optical detection system is designed to provide stable absorbance measurements across the specified wavelength range. The photometric range reaches 0–4.0 OD, with a resolution of 0.001 OD.
These characteristics make the SpectraMax 190 suitable for quantitative biochemical analysis and routine laboratory measurements.
PathCheck technology is one of the distinguishing features of the SpectraMax 190.
In conventional microplate measurements, different liquid volumes can produce different optical pathlengths, affecting absorbance readings.
PathCheck automatically compensates for liquid pathlength differences and helps convert microplate measurements into equivalent 1 cm cuvette values.
This function is particularly useful for DNA, RNA, and protein concentration measurements.
The SpectraMax 190 is optimized for 96-well microplate analysis.
It supports endpoint and kinetic reading modes, allowing researchers to perform both single-time-point measurements and time-dependent experiments.
The endpoint reading time is approximately 12 seconds for a standard 96-well plate, depending on the measurement configuration.
The instrument provides temperature regulation from ambient temperature +4°C to 45°C.
Temperature uniformity is approximately ±0.5°C at 37°C across the microplate wells.
This function supports temperature-sensitive biochemical assays and enzyme kinetic experiments.
The following table summarizes the main technical specifications of the Molecular Devices SpectraMax 190.
|
Parameter |
Specification |
|---|---|
|
Manufacturer |
Molecular Devices |
|
Model |
SpectraMax 190 |
|
Product Type |
UV-Visible Absorbance Microplate Reader |
|
Detection Method |
Absorbance |
|
Light Source |
Xenon Flash Lamp |
|
Wavelength Range |
190–850 nm |
|
Wavelength Selection |
Monochromator |
|
Wavelength Increment |
1 nm |
|
Wavelength Bandwidth |
2 nm |
|
Wavelength Accuracy |
±1.0 nm |
|
Wavelength Repeatability |
±0.2 nm |
|
Photometric Range |
0–4.0 OD |
|
Photometric Resolution |
0.001 OD |
|
Photometric Accuracy |
±0.006 OD ±1.0%, 0–2.0 OD |
|
Photometric Precision |
±0.003 OD ±1.0%, 0–2.0 OD |
|
Stray Light |
≤0.05% at 230 nm |
|
Microplate Format |
96-Well |
|
Reading Modes |
Endpoint and Kinetic |
|
Endpoint Reading Time |
Approx. 12 seconds |
|
Kinetic Reading Interval |
Minimum 9 seconds |
|
Temperature Range |
Ambient +4°C to 45°C |
|
Temperature Uniformity |
±0.5°C at 37°C |
|
PathCheck Technology |
Yes |
|
Software |
SoftMax Pro |
|
Dimensions (W × D × H) |
58 × 38 × 22 cm |
|
Weight |
Approx. 13.6 kg |
|
Power Requirements |
90–250 VAC, 50/60 Hz |
|
Power Consumption |
Less than 250 W |
Specifications are based on the manufacturer's SpectraMax 190 product datasheet and operating documentation. Actual configuration and software version should be verified on the individual instrument.
The Molecular Devices SpectraMax 190 is suitable for various absorbance-based laboratory applications.
DNA and RNA Quantification
The instrument can measure nucleic acid absorbance at 260 nm and support DNA and RNA concentration analysis. Measurements at 280 nm can also help assess protein contamination.
Protein Concentration Analysis
Researchers can use the SpectraMax 190 for colorimetric protein assays, including Bradford and BCA methods, depending on the selected assay protocol.
ELISA and Immunoassays
The instrument supports absorbance-based ELISA applications and other immunoassays. Its 96-well format makes it suitable for routine sample screening and biochemical analysis.
Enzyme Activity Measurements
Its kinetic reading function enables researchers to monitor absorbance changes over time, making it useful for enzyme activity studies and biochemical reaction analysis.
Pharmaceutical Research
Pharmaceutical laboratories can use the SpectraMax 190 for dissolution-related absorbance measurements, colorimetric assays, and other UV-visible analytical procedures.
Proper maintenance is essential for ensuring long-term instrument performance and reliable analytical results.
Optical System Inspection: Regularly check the optical measurement system and xenon flash lamp performance.
Microplate Chamber Cleaning: Keep the reading chamber and plate carrier clean to prevent contamination.
Temperature Verification: Periodically verify temperature regulation and uniformity when temperature-controlled assays are performed.
PathCheck Verification: Check PathCheck performance according to the manufacturer's recommended procedures.
Software Maintenance: Maintain compatible SoftMax Pro software and ensure reliable computer communication.
Performance Validation: Periodically verify wavelength accuracy, photometric accuracy, and measurement repeatability.
For used instruments, additional inspection of the lamp, optical system, plate transport mechanism, temperature control, and software functionality is recommended before purchase.
The Molecular Devices SpectraMax 190 Microplate Reader combines flexible UV-visible wavelength selection, reliable absorbance measurement, PathCheck technology, and convenient 96-well microplate processing.
Its broad wavelength range and compatibility with common biochemical assays make it a valuable instrument for molecular biology, biotechnology, pharmaceutical research, and academic laboratories.
For laboratories seeking cost-effective equipment solutions, a professionally inspected and refurbished SpectraMax 190 can be an economical alternative to purchasing a new microplate reader.
Before purchasing a used instrument, buyers should confirm its optical performance, lamp condition, temperature regulation, plate carrier operation, PathCheck functionality, and software compatibility.
Molecular Devices SpectraMax 190 – Accurate UV-Visible Absorbance Analysis for Reliable Laboratory Research.
Product Inquiry